Tips_3000 cleaning

gonotec osmomat 3000 cleaning 200x2004

Osmomat 3000 Cleaning

GOCLEAN0009Be sure unit power supply is disconnected! Risk of electric shock!

GOCLEAN0009Unit has been decontaminated especially the initiation needle ! Risk of injury from the tip of the needle!

GOCLEAN0009Use appropriate protective gear.

GOCLEAN0010WARNING! Risk of infection from sample residue!

GOCLEAN0020


1 Touchscreen.
2 Upper cooling system (behind movable elevator cover).
3 Temperature sensor with measurement vessel.
4 Lower cooling system.
5 Elevator.
6 Printer (Option D)(does not apply to OSMOMAT 3000basic).


Needle cleaning

Clean the needle at least once every quarter. Tools are needed (Allen key SW 2) Power down device using on/off switch on rear side and disconnect power plug. For cleaning the needle remove the elevator cover with the 2 screws on the front as shown in (figure 1) and pull carefully the cover off.

Figure 1: Elevator cover

GOCLEAN0021

 

 

 

 

 

 

Remove device front panel by pulling forward (figure 2).The device front panel is secured to the housing by magnets.


GOCLEAN0022

Figure 2: Front panel


Remove mounting screw (Figure 3) of cryst-needle while holding cryst-needle in place bring it upwards.
Pay attention to the needle guide don’t forget to place it back otherwise the needle will bend and get damage (Risk of injury from the tip of the needle!)


GOCLEAN0023


Figure 3: Remove mounting screw

GOCLEAN0024



 

 

1 Needle guide tube
2 needle
3 Screw and ringUse a soft paper towel to thoroughly clean the thermistor probe you can moistened the paper with water or about 50% isopropanol.


The needle may have a slight bend towards the back

GOCLEAN0025Figure 4: Straighten cryst-needle



 

 

 

 

 

When placing the front cover and elevator cover back pay attention to the cable’s.

GOCLEAN0026

Figure 5: cover placement


Install device front panel on device and push into place (figure 6).

GOCLEAN0027

Figure 6: front cover


Thermistor probe and Upper Peltier element cleaning

Clean upper peltier element and probe after each batch or at the end of the day this will prevent de-icing of the element.


Power down device using on/off switch on rear side and disconnect power plug.


For cleaning the thermistor probe use a soft paper towel to thoroughly clean the thermistor probe you can moistened the paper with water ore.


For cleaning the upper peltier element use a cotton swab stick drained it in 50% isopropanol or water depending on the pollution to clean the peltier element.

Wipe off dirt and dust from housing using a soft cloth.

GOCLEAN0028

 

 

 

 

 

 

 

Figure 7: swab stick/ Q-tips


Lower Peltier element

Clean lower peltier element after each batch or at the end of the day.
Use a cotton swab stick drained in 50% isopropanol or water depending on the pollution to clean the peltier element.

GOCLEAN0029