Tips_Option Reset timer

Tips_Option Reset timer

Reset Replacement Timer

Turn the unit off at the power swich.

To enter the replacement timer set-up, press the left hand touch pad button on the control panel and at the same time turn the power on.

ELREPLACE0001

 

 

 

 

 

 

 

 

 

 


ELGAFILTER0002CAUTION! Before re-setting any of the cartridge timers, ensure that the appropriate new cartridges have been installed and securely located correctly in the Elga PURELAB


Step 1 - Ion-exchange Cartridge Timer
ELREPLACE0002



 

 

 

 

Setting this screen will cause the ion-exchange cartridge timer to reset to the preset value of 4380 hours (6 months).
Press ELREPLACE0003to jump to the next consumable or ELREPLACE0004 to initiate reset.
Press ELREPLACE0003 to reset timer or press ELREPLACE0005 to abort reset.
Press ELREPLACE0003 to jump to the next consumable.



Ion-exchange Cartridge Timer


Step 2 - UV Lamp Replacement Timer

Setting this screen will cause the UV Lamp timer to reset to the preset value of 8760 hours (1 year).
Press ELREPLACE0003to jump to the next consumable or ELREPLACE0004 to initiate reset.
Press ELREPLACE0003to reset timer or press ELREPLACE0005 to abort reset.
Press ELREPLACE0003to jump to the next consumable.

ELREPLACE0008

UV Lamp Replacement Timer

 

 

 

 

 

 

 

 

Step 3 - Pre-treatment Cartridge Replacement Timer

Setting this screen will cause the pre-treatment cartridge timer to reset to the preset value of 4380 hours (6 months).
PressELREPLACE0003to jump to the next consumable or ELREPLACE0004 to initiate reset.
Press ELREPLACE0003to reset timer or press ELREPLACE0005 to abort reset.
PressELREPLACE0003 to jump to the next consumable.


Pre-treatment Cartridge Timer


Identification of Consumables
There are three types of unique replacement consumables designed for use in the PURELAB Option units and these are illustrated with the following part numbers:
LC140 Pre-treatment cartridge
LC141 Ion-exchange cartridge
LC105 UV lamp

Frequency of Consumable Replacement
The following frequency of consumable replacement is recommended as a guide assuming typical usage*:

Pre-treatment - LC140 max 6 months
Ion-exchange - LC141 max 6 months
UV lamp - LC105 max 12 months
Reverse Osmosis - LC143 every 2 - 3 years (not an operator replacement item)

These frequencies are only estimates and replacement will depend on the application and feed water quality.