Reset Replacement Timer
Turn the unit off at the power swich.
To enter the replacement timer set-up, press the left hand touch pad button on the control panel and at the same time turn the power on.
CAUTION! Before re-setting any of the cartridge timers, ensure that the appropriate new cartridges have been installed and securely located correctly in the Elga PURELAB
Step 1 - Ion-exchange Cartridge Timer
Setting this screen will cause the ion-exchange cartridge timer to reset to the preset value of 4380 hours (6 months).
Press to jump to the next consumable or to initiate reset.
Press to reset timer or press to abort reset.
Press to jump to the next consumable.
Ion-exchange Cartridge Timer
Step 2 - UV Lamp Replacement Timer
Setting this screen will cause the UV Lamp timer to reset to the preset value of 8760 hours (1 year).
Press to jump to the next consumable or to initiate reset.
Press to reset timer or press to abort reset.
Press to jump to the next consumable.
UV Lamp Replacement Timer
Step 3 - Pre-treatment Cartridge Replacement Timer
Setting this screen will cause the pre-treatment cartridge timer to reset to the preset value of 4380 hours (6 months).
Pressto jump to the next consumable or to initiate reset.
Press to reset timer or press to abort reset.
Press to jump to the next consumable.
Pre-treatment Cartridge Timer
Identification of Consumables
There are three types of unique replacement consumables designed for use in the PURELAB Option units and these are illustrated with the following part numbers:
LC140 Pre-treatment cartridge
LC141 Ion-exchange cartridge
LC105 UV lamp
Frequency of Consumable Replacement
The following frequency of consumable replacement is recommended as a guide assuming typical usage*:
Pre-treatment - LC140 max 6 months
Ion-exchange - LC141 max 6 months
UV lamp - LC105 max 12 months
Reverse Osmosis - LC143 every 2 - 3 years (not an operator replacement item)
These frequencies are only estimates and replacement will depend on the application and feed water quality.